Philips Issues Urgent Correction for Transportable Telemetry System-2 (MCOT) Due to ECG Data Routing Error

Philips has initiated an urgent medical device correction for its Monitoring Service Application, associated with the Mobile Cardiac Outpatient Telemetry (MCOT) system, also referred to as Transportable Telemetry System-2. This correction, classified by the FDA as a Class I recall – the most serious type – addresses a critical software configuration issue that could lead to serious health consequences or death if left unaddressed. This recall emphasizes the importance of accurate and timely cardiac monitoring and reporting in patient care. It’s crucial to understand that this is a correction, not a removal of devices from service, but requires immediate action from healthcare providers to ensure patient safety.

Understanding the Issue with Philips Transportable Telemetry System-2 (MCOT)

The core of the problem lies within the Monitoring Service Application of the Philips MCOT, a transportable telemetry system used for cardiac monitoring. Between July 2022 and July 2024, a software error caused some electrocardiogram (ECG) events to be improperly routed within the system. This misrouting meant that these critical cardiac events were not reviewed by cardiology technicians as intended, potentially leading to a breakdown in the reporting process to clinicians.

Specifically, ECG events indicative of serious arrhythmias, such as atrial fibrillation, pauses, supraventricular tachycardia, ventricular tachycardia, and second or third-degree AV block, were among those potentially missed. These are conditions requiring prompt medical attention, and failure to identify them in a timely manner can have significant repercussions for patient health.

Potential Impact on Patients and Clinical Decision Making

The consequence of this software malfunction is potentially severe. Missed or delayed review of critical ECG events can lead to:

  • Delayed Diagnosis: Conditions like atrial fibrillation or ventricular tachycardia might go undetected for longer periods.
  • Inadequate Treatment: Without timely information, healthcare providers may not be able to initiate or adjust treatment plans promptly.
  • Serious Adverse Health Events: Prolonged periods of untreated arrhythmias can result in serious health consequences, including injury and, in the most severe cases, death.

The FDA reports already being aware of 109 injuries and 2 deaths potentially linked to this issue, highlighting the urgency and seriousness of this correction. The accuracy of cardiac monitoring and reporting from systems like the transportable telemetry system-2 is paramount for effective clinical decision-making and patient safety.

Philips’ Recommended Actions for Healthcare Providers

In response to this critical issue, Philips issued an Urgent Medical Device Correction notice on December 18, 2024, to all affected customers. Healthcare providers using the Philips MCOT system, a key example of a transportable telemetry system-2, are directed to take the following immediate steps:

  1. Review Frequently Asked Questions (FAQs): Philips has provided a detailed FAQ document to address common concerns and questions related to this correction. This document is a vital resource for understanding the issue and the necessary steps.
  2. Access the Prescriber Response Site: Using the unique Location Code provided in the notification letter, log into the secure Prescriber Response Site at https://prs.gobio.com.
  3. Acknowledge Receipt of the Urgent Correction Letter: Confirm receipt and understanding of the correction notice directly on the Prescriber Response Site. This step is crucial for Philips to track awareness and ensure all affected users are informed.
  4. Review Patient List and Select Reprocessing Options: The Prescriber Response Site contains a list of patients potentially impacted by this software issue. Healthcare providers must review this list and choose one of the following reprocessing options:
    • Reprocess All: Request reprocessing of ECG data for all listed patients.
    • Reprocess None: Decline reprocessing for all listed patients (this option should be carefully considered given the potential risks).
    • Selected Patients: Choose specific patients from the list for ECG data reprocessing.

Philips emphasizes that they will not be contacting patients directly. It is the responsibility of healthcare providers to inform patients who may have been affected and to manage their care pathways accordingly.

Frequently Asked Questions for Clinicians

To further assist healthcare providers, Philips has provided answers to key questions regarding this transportable telemetry system-2 correction:

How can I identify my impacted patients?

A comprehensive list of potentially affected patients is available for review within the secure Prescriber Response Site.

What is the Prescriber Response Site and its purpose?

The Prescriber Response Site (https://prs.gobio.com) is a secure online portal specifically created for this device correction. It serves multiple purposes:

  • Providing access to the list of your potentially impacted patients.
  • Hosting reprocessed ECG reports in the form of Delta Summary Reports.
  • Delivering notifications if reprocessing reveals no changes to the original reports.

What specific actions are required from my clinical practice?

The necessary actions involve utilizing the Prescriber Response Site:

  • Log in using the provided Location Code and your NPI (National Provider Identifier).
  • Acknowledge receipt of the Urgent Medical Device Correction Letter.
  • Review the patient list and select your preferred reprocessing option (All, None, or Selected patients).

What will Philips do if I choose to generate Delta Summary Reports?

If you opt for reprocessing and Delta Summary Reports, Philips will:

  • Generate these reports for the patients you selected.
  • Make the reports available for review on the Prescriber Response Site.
  • Notify you via email (using the address provided during acknowledgement) when the reports are ready for your review.

How will newly identified Urgent and Emergent events be handled during reprocessing?

Any newly identified urgent or emergent events discovered during the reprocessing will be included in the Delta Summary Report. However, it is important to note that these newly identified critical events will not be communicated through routine notification channels. Clinicians must actively review the Delta Summary Reports on the Prescriber Response Site.

Who should I contact if I encounter adverse reactions or quality issues related to this software problem?

For any further information, support, or to report adverse reactions or quality problems associated with this software configuration issue, please contact Philips’ Prescriber Response Line at 888-521-1684.

Importance of Unique Device Identification (UDI)

This recall also highlights the importance of the Unique Device Identifier (UDI) system. The UDI helps to accurately track medical devices, like this transportable telemetry system-2, throughout their lifecycle, from manufacturing to patient use. This enhanced traceability is crucial for:

  • Accurate Reporting: Facilitating more precise reporting of adverse events.
  • Efficient Review and Analysis: Improving the FDA’s ability to review and analyze adverse event reports.
  • Faster Problem Correction: Enabling quicker identification of device-related problems and more efficient implementation of corrections and recalls, ultimately contributing to improved patient safety.

Reporting Problems to the FDA

Beyond contacting Philips directly, healthcare professionals and patients are also encouraged to report any adverse reactions or quality problems experienced while using these devices to MedWatch, the FDA’s Safety Information and Adverse Event Reporting Program. Reports can be submitted online at https://www.accessdata.fda.gov/scripts/medwatch/index.cfm?action=reporting.home.

This situation underscores the critical role of reliable transportable telemetry systems like the Philips MCOT in modern healthcare and the vital importance of prompt action in response to device recalls and corrections to safeguard patient well-being.